Electricity, electrical safety and diathermy
The basic electrical concepts behind monitoring and theatre equipment (current, voltage, resistance, impedance, capacitance and inductance), how the ECG and other bioelectric signals are recorded, the effects of current on the body, macroshock and microshock, how earthing, isolation, residual current devices and equipment classes protect patients, and the principles and hazards of surgical diathermy and the defibrillator.
By the end of this page you should be able to
- Define current, potential difference, resistance, impedance, capacitance and inductance, and apply Ohm's law.
- Explain how bioelectric signals such as the ECG are recorded and how interference is rejected.
- Describe the effects of electric current on the body and distinguish macroshock from microshock.
- Describe how earthing, isolated supplies, line isolation monitors, residual current devices and equipment classification protect patients.
- Explain the principles of monopolar and bipolar diathermy and of the defibrillator, and their hazards.
Basic concepts
- Current (amperes) is the rate of flow of charge; one ampere is one coulomb per second. Potential difference (volts) is the energy per unit charge (joules per coulomb) that drives current. Resistance (ohms) opposes direct current. Ohm's law: V = I × R. Power = V × I = I² × R, so heating rises with the square of current.
- Direct current flows one way; alternating current reverses periodically (mains supply in Australia and New Zealand is 230 V at 50 Hz).
- Capacitance is the ability to store charge, measured in farads; a capacitor is two conductors separated by an insulator. It blocks direct current but passes alternating current, and its opposition to current (reactance) falls as frequency rises. Capacitance between equipment and the patient is how leakage current flows even without a direct connection. The defibrillator stores its energy in a capacitor (energy = ½ × C × V²).
- Inductance is the opposition of a coil to a change in current, from the magnetic field it generates; its reactance rises with frequency. Inductors are used to shape the defibrillator waveform and in transformers.
- Impedance is the total opposition to alternating current, combining resistance and reactance; it depends on frequency. Skin impedance is high when dry and much lower when wet or broken.
- Transformers use two coils on a shared core; alternating current in the primary induces current in the secondary with no direct electrical connection, which is the basis of isolation.
Recording bioelectric signals
The ECG records the potential differences on the body surface generated by cardiac depolarisation and repolarisation, about 1 mV in amplitude with useful frequencies of roughly 0.05 to 150 Hz for diagnosis (a narrower band of about 0.5 to 40 Hz is used in monitoring mode to reduce artefact). The EEG is smaller still (microvolts), and the EMG has higher frequencies.
Such small signals are vulnerable to interference: mains hum at 50 Hz induced capacitively or inductively from nearby cables, diathermy, movement, and muscle activity. The defences are silver/silver chloride electrodes with conductive gel to reduce and stabilise skin impedance; screened leads; a differential amplifier that amplifies the difference between two inputs and rejects signals common to both (expressed as the common mode rejection ratio, ideally very high), so interference that appears on both leads equally cancels; and filters (a notch filter at mains frequency, high-pass and low-pass filters). Filtering trades fidelity for stability: a monitoring-mode filter can distort the ST segment.
Electric shock
Harm depends on the current that flows (not the voltage alone), its path (across the heart is worst), duration, frequency (50 to 60 Hz is the most dangerous; high frequencies above about 100 kHz do not depolarise excitable tissue) and the current density at any point. With mains current applied across the skin from hand to hand or hand to foot, roughly: 1 mA is perceived, 5 mA is painful, 10 to 20 mA causes sustained muscle contraction so the victim cannot let go, about 50 mA affects respiratory muscles, and about 100 mA or more can cause ventricular fibrillation. This is macroshock. Larger currents cause burns, which follow I² × R heating at the points of entry and exit and along the path, including deep tissues.
Microshock occurs when current is delivered directly to the heart, through a pacing wire, a saline-filled central catheter or a pulmonary artery catheter. Because the current density at the myocardium is then high, as little as about 100 µA (0.1 mA) can cause fibrillation, well below the threshold of perception, so neither the patient nor the staff will notice it. This is why equipment that connects to the heart must meet much stricter leakage current limits.
Protecting the patient
- Earthing: the exposed metal case of mains equipment is connected to earth, so a fault that makes the case live sends a large current to earth and blows the fuse rather than passing through a person. The patient should not be earthed, because an earthed patient completes a circuit for any stray current.
- Isolated (floating) supplies: an isolating transformer supplies the theatre or the patient-connected part with no reference to earth, so touching a single live conductor while earthed does not complete a circuit. A line isolation monitor warns when isolation has been lost.
- Residual current devices (core balance or earth leakage breakers) compare current in the active and neutral conductors and cut the supply within milliseconds when they differ by more than a small threshold, as they will if current is leaking to earth through a person. Body-protected areas use them to meet the cardiac-protected and body-protected standards.
- Equipment classes: Class I has an earthed case; Class II has double or reinforced insulation and needs no earth; Class III runs on safety extra-low voltage from an isolated source.
- Applied part types: B (body), BF (body, floating: the patient connection is isolated) and CF (cardiac, floating: isolated and with the lowest permitted leakage current, about 10 µA in normal conditions), which is required for parts with direct cardiac connection.
- Other measures: equipotential earthing so that all metal near the patient is at the same potential, regular testing of equipment, avoiding ungrounded multiple outlet boards, keeping fluids away from mains equipment, and never touching a pacing wire and a mains device at the same time.
Diathermy and the defibrillator
Surgical diathermy passes high-frequency alternating current (around 0.5 to 1 MHz) through tissue. At that frequency current does not depolarise nerve and muscle, so it heats tissue without causing a shock. Heating depends on current density: in monopolar diathermy, current flows from a small active electrode (high density: cutting or coagulation) through the patient to a large return plate (low density: no heating) and back to the generator. A continuous sine wave cuts by vaporising cells; an interrupted, higher-voltage waveform coagulates. Bipolar diathermy passes current only between the tips of the forceps, so it needs no plate, uses less power and is safer near pacemakers and in end organs with narrow pedicles.
Hazards: burns at the plate if contact is poor or the plate is partly detached (current density rises); burns at alternative earth points (ECG electrodes, metal contact) with older earth-referenced machines; current channelling along narrow structures (a penis or digit); interference with or reprogramming of pacemakers and defibrillators, and inhibition of pacing; ignition of alcohol skin preparation, bowel gas or oxygen-enriched atmospheres in the airway; and surgical smoke. Modern generators are isolated and monitor plate contact.
The defibrillator charges a capacitor from a direct current source and discharges it across the chest, depolarising a critical mass of myocardium so that organised activity can resume. Biphasic waveforms reverse current part way through, achieving defibrillation at lower energy with less myocardial injury, and modern devices compensate for transthoracic impedance. Synchronised cardioversion times the shock to the R wave to avoid the vulnerable period on the T wave.
Pulling it together
- Define each quantity with its unit and a clinical example before using it.
- For shock, give the determinants (current, path, duration, frequency, density) before the thresholds.
- Explain microshock by current density, and link it to CF equipment and handling of pacing wires.
- Explain why diathermy at high frequency heats without shocking, and why the plate must be large and in good contact.
Common traps
- Saying voltage alone determines harm; current and its density do.
- Saying the patient should be earthed for safety; the case is earthed, the patient is not.
- Confusing a fuse (protects the equipment and wiring from overload) with a residual current device (protects people from leakage).
- Saying capacitive reactance rises with frequency; it falls.
- Thinking bipolar diathermy needs a return plate.
Where this meets intensive care
- Temporary epicardial pacing wires and saline-filled central lines give current a direct path to the heart, so microshock levels far below perception can cause fibrillation; insulate exposed wire ends and handle them with gloves.
- A wandering or noisy ECG baseline in ICU often comes from poor electrode contact or nearby equipment; good skin preparation and electrode placement fix more than filters do.
Viva questions
Answer each one aloud, as you would to an examiner, before you open the model points.
- 1.
Define capacitance and inductance, and give a clinical use of each.
Show answer
- Capacitance: ability to store charge, farads; blocks direct and passes alternating current, reactance falls with frequency.
- Use: defibrillator energy storage; leakage current through stray capacitance.
- Inductance: opposition to change in current from a self-induced magnetic field, henries; reactance rises with frequency.
- Use: shaping the defibrillator discharge, transformers, filters.
- 2.
How is the ECG recorded without being swamped by interference?
Show answer
- Signal about 1 mV; interference from mains, diathermy, movement and muscle.
- Silver/silver chloride electrodes and gel to lower skin impedance; screened leads.
- Differential amplifier with a high common mode rejection ratio.
- Filters, with a monitoring mode narrower than diagnostic mode, at the cost of ST fidelity.
- 3.
What factors determine the harm from an electric shock?
Show answer
- Current magnitude, which depends on voltage and impedance (especially skin).
- Path across the body, current density at the heart.
- Duration and frequency: 50 to 60 Hz most dangerous; high frequencies do not depolarise.
- Thresholds for perception, pain, tetany, respiratory effects and fibrillation; burns from I²R heating.
- 4.
What is microshock and how is it prevented?
Show answer
- Current delivered directly to the myocardium through a catheter or pacing wire.
- Very high current density, so about 100 µA can cause fibrillation, below perception.
- CF-rated equipment with very low leakage current; isolated and equipotential systems.
- Handle pacing wires with gloves, insulate exposed terminals, avoid touching wires and equipment together.
- 5.
How does a residual current device work?
Show answer
- Active and neutral conductors pass through a toroidal core.
- Normally currents are equal and opposite, so no net magnetic flux.
- Leakage to earth through a person creates an imbalance, induces current in a sensing coil and trips the supply within milliseconds.
- Protects people, unlike a fuse, which protects against overload.
- 6.
Explain the equipment classes and applied part types.
Show answer
- Class I: earthed case with fuses; Class II: double insulation; Class III: safety extra-low voltage.
- Type B: applied part not isolated; BF: floating, isolated from earth.
- Type CF: isolated with the lowest leakage limits, required for direct cardiac connection.
- Choice depends on how and where the device touches the patient.
- 7.
Why does diathermy not electrocute the patient?
Show answer
- High frequency current (around 0.5 to 1 MHz) changes direction too quickly to depolarise excitable membranes.
- Heating depends on current density: high at the small active tip, low at the large return plate.
- Cutting uses a continuous waveform; coagulation an interrupted, higher voltage waveform.
- Bipolar confines current between the forceps tips.
- 8.
What are the hazards of diathermy, and how are they reduced?
Show answer
- Plate burns from poor contact or partial detachment; burns at alternative earth points with older machines.
- Current channelling along narrow structures; capacitive coupling in laparoscopic instruments.
- Pacemaker and defibrillator interference; fires with alcohol preparations or oxygen-rich atmospheres; smoke.
- Isolated generators with plate monitoring, correct plate siting, bipolar where possible, device reprogramming as indicated.
- 9.
How does a defibrillator work?
Show answer
- Direct current charges a capacitor; energy stored = ½ C V².
- Discharge across the chest depolarises a critical mass of myocardium.
- Biphasic waveforms work at lower energy with less injury; impedance compensation.
- Synchronised mode times the shock to the R wave to avoid the vulnerable period.
Practise this week
- Investigations Lab: gases, ECGs and lung function
- First Part MCQs: measurement and equipment
- First Part SAQs, marked part by part
- Anatomy Scroller: procedural anatomy on a real CT
Week 7 in the other subjects
- Physiology: Cardiac electrophysiology and the ECG
- Pharmacology: Antiarrhythmic drugs
Not affiliated with the CICM. Notes are a revision aid, not a substitute for the curriculum or for clinical guidelines.