Gas laws, flow and oxygen delivery
The gas laws and the ideas of partial pressure, critical temperature and vapour, laminar and turbulent flow and the Reynolds number, the Bernoulli principle and entrainment, how flow is measured, how oxygen is stored and supplied, and why variable and fixed performance devices and high-flow nasal oxygen behave as they do.
By the end of this page you should be able to
- State and apply Boyle's, Charles' and Gay-Lussac's laws, the universal gas law, Dalton's law and Henry's law.
- Define critical temperature and explain why oxygen is stored as a compressed gas and nitrous oxide as a liquid.
- Describe laminar and turbulent flow, the Hagen-Poiseuille relationship and the Reynolds number, and apply them to airways and cannulae.
- Explain the Bernoulli principle and the Venturi effect and how they are used in oxygen delivery.
- Compare variable and fixed performance oxygen devices and explain how high-flow nasal oxygen works.
The gas laws
- Boyle's law: at constant temperature, pressure × volume is constant. A cylinder's contents can be estimated from its pressure for a gas stored above its critical temperature, and gas in a closed space expands as ambient pressure falls (air in the middle ear, a pneumothorax or a tube cuff at altitude).
- Charles' law: at constant pressure, volume is proportional to absolute temperature. Gay-Lussac's law (the third gas law): at constant volume, pressure is proportional to absolute temperature, which is why a cylinder's pressure rises when it is heated.
- Universal gas law: PV = nRT, combining the three with Avogadro's hypothesis that equal volumes of gases at the same temperature and pressure contain equal numbers of molecules; one mole occupies about 22.4 L at standard temperature and pressure.
- Dalton's law: in a mixture, each gas exerts the pressure it would exert alone, and the total is the sum of these partial pressures. Partial pressure = fraction × total pressure, which is why inspired oxygen tension falls at altitude though the fraction is unchanged, and why the inspired value must first subtract the saturated vapour pressure of water (47 mmHg at 37 °C).
- Henry's law: at a given temperature, the amount of gas dissolved in a liquid is proportional to its partial pressure above it. Dissolved oxygen is 0.003 mL/100 mL per mmHg, which is why breathing pure oxygen adds only a little to content, and why hyperbaric oxygen can carry meaningful dissolved oxygen.
- Graham's law: the rate of diffusion of a gas is inversely proportional to the square root of its molecular weight; Fick's law governs diffusion across a membrane.
Critical temperature, vapour and storage
A gas is a substance above its critical temperature, the temperature above which it cannot be liquefied however much pressure is applied. A vapour is the gaseous form of a substance below its critical temperature. Oxygen's critical temperature is about minus 119 °C, so at room temperature it is a gas and is stored compressed in cylinders (a full cylinder is filled to well over 100 bar, the exact figure depending on the cylinder type); the pressure falls in proportion to the contents, so the gauge reads the amount left. Large hospitals store oxygen as a cold liquid in a vacuum-insulated evaporator, where liquid boils off as needed and is piped at about 400 kPa.
Nitrous oxide has a critical temperature of about 36.5 °C, so at room temperature it is stored as a liquid in equilibrium with its vapour. The gauge shows the saturated vapour pressure (about 44 to 52 bar at room temperature), which stays constant until all the liquid has gone; the contents must be found by weighing the cylinder. A filling ratio (mass of nitrous oxide ÷ mass of water the cylinder could hold) limits filling so that a rise in temperature cannot burst it. The pseudocritical temperature explains why the 50:50 oxygen and nitrous oxide mixture can separate (laminate) if cooled below about minus 6 °C, delivering high oxygen first and then hypoxic nitrous oxide.
When a compressed gas expands rapidly it cools (the Joule-Thomson effect), which is why a cylinder valve can frost. When a liquid evaporates it takes latent heat from its surroundings, which is why nitrous oxide cylinders cool and frost during heavy use and the vapour pressure falls.
Flow
In laminar flow, fluid moves in smooth parallel layers, fastest in the centre with a parabolic profile. The Hagen-Poiseuille relationship gives flow = (π × pressure difference × radius⁴) ÷ (8 × viscosity × length). Flow is therefore proportional to the pressure gradient and to the fourth power of the radius, and inversely proportional to viscosity and length. Halving a tube's radius cuts flow sixteenfold at the same pressure: hence wide, short cannulae for resuscitation, and the large rise in airway resistance with a small amount of bronchoconstriction or secretions.
In turbulent flow, eddies form, the profile is flattened, and much more energy is lost. Flow becomes proportional to the square root of the pressure gradient, so doubling flow needs about four times the pressure, and it depends on density rather than viscosity. Whether flow is laminar or turbulent is predicted by the Reynolds number = (density × velocity × diameter) ÷ viscosity; above about 2000 turbulence becomes likely. Turbulence is promoted by high flow, sharp bends, branches, constrictions and rough walls. In the lungs, flow is turbulent in the trachea and large bronchi, transitional in the middle airways and laminar in the small airways, where total cross-sectional area is huge and velocity low. Heliox (helium and oxygen) has a much lower density than air, lowering the Reynolds number and the pressure needed to drive flow through a large-airway obstruction, at the cost of limiting the oxygen fraction.
Bernoulli: along a streamline, the sum of pressure energy and kinetic energy is constant, so where a tube narrows and velocity rises, lateral pressure falls. The Venturi effect uses this: a jet of oxygen through a narrow orifice creates a low-pressure region that entrains room air through side ports in a fixed ratio, producing a predictable oxygen fraction. Nebulisers, suction devices and jet ventilation injectors use the same principle. The Coanda effect describes a fluid jet attaching to a nearby curved surface, which can make flow favour one branch of a bifurcation.
Measuring flow
- Variable orifice, constant pressure (rotameter): a bobbin rises in a tapered tube until the pressure drop across the annular gap balances its weight. At low flow the gap is long and narrow, so flow is laminar and depends on viscosity; at high flow the gap is short and wide, so it acts as an orifice and depends on density. Each is calibrated for one gas; errors arise from static, dirt, tilting and back pressure.
- Fixed orifice, variable pressure: a pneumotachograph measures the pressure drop across a fixed resistance of laminar elements, proportional to flow; it must be heated to prevent condensation and calibrated for the gas mixture. Orifice and pitot devices measure pressure drop under turbulent conditions (proportional to flow squared).
- Others: hot-wire anemometers (cooling of a heated wire), ultrasonic flowmeters (transit time or Doppler shift), and the Wright respirometer (a vane turned by flow, reading high at high flow and low at low flow because of inertia).
Oxygen delivery devices
A patient's peak inspiratory flow is about 30 L/min at rest and may exceed 60 to 100 L/min in respiratory distress. Whether a device delivers a predictable oxygen fraction depends on whether it supplies more than that flow.
Variable performance devices (nasal cannulae, simple face masks, partial rebreathing masks) supply less than the peak inspiratory flow, so the patient entrains room air and the inspired fraction varies with the breathing pattern: the faster and deeper the breathing, the lower the fraction. Nasal cannulae use the nasopharynx as a small reservoir; a simple mask uses the mask volume and needs enough flow to flush exhaled carbon dioxide. A non-rebreathing mask with reservoir gives a high but still variable fraction.
Fixed performance devices supply flow at or above peak inspiratory flow at a known fraction. Venturi masks do this by entrainment: the lower the set oxygen fraction, the more air is entrained and the higher the total flow. A sealed anaesthetic circuit or a ventilator is also fixed performance.
High-flow nasal oxygen delivers heated, fully humidified gas at up to about 60 L/min through wide cannulae. It works by meeting or exceeding inspiratory flow so the fraction is reliable; washing out nasopharyngeal dead space so each breath carries less carbon dioxide; generating a small, flow-dependent positive pressure (a few cmH2O with the mouth closed) that helps recruit alveoli; and reducing the work of conditioning cold dry gas, with better secretion clearance and comfort.
Answering like a physicist
- State each law precisely with its constant variables, then give one clinical application.
- Define gas and vapour by critical temperature before talking about cylinders.
- For flow, write the Hagen-Poiseuille relationship, then contrast turbulent flow and give the Reynolds number with its components.
- Classify oxygen devices by performance (fixed against variable) and explain the deciding principle: device flow against patient peak inspiratory flow.
Common traps
- Using the gauge pressure to estimate the contents of a nitrous oxide cylinder.
- Saying turbulent flow depends on viscosity; it depends on density.
- Saying a Venturi increases pressure at the constriction; pressure falls as velocity rises.
- Quoting an inspired oxygen fraction for nasal cannulae as if it were fixed.
- Forgetting water vapour pressure when calculating inspired oxygen tension.
Where this meets intensive care
- High-flow nasal oxygen gives a predictable oxygen fraction only when its flow meets the patient's peak inspiratory flow; in severe distress that demand can exceed what the device delivers.
- A narrowed airway or small tracheostomy tube raises resistance steeply because flow depends on the fourth power of the radius when laminar and on density when turbulent.
Viva questions
Answer each one aloud, as you would to an examiner, before you open the model points.
- 1.
What is the difference between a gas and a vapour, and why does it matter for cylinders?
Show answer
- A gas is above its critical temperature and cannot be liquefied by pressure; a vapour is below it.
- Oxygen (critical temperature about minus 119 °C) is a compressed gas; gauge pressure falls in proportion to contents.
- Nitrous oxide (about 36.5 °C) is a liquid in equilibrium with vapour; gauge pressure is constant until the liquid is gone, so contents are found by weight.
- Filling ratio limits nitrous oxide filling to allow for thermal expansion.
- 2.
State Dalton's and Henry's laws and apply them.
Show answer
- Dalton: total pressure is the sum of partial pressures; partial pressure equals fraction times total pressure.
- Application: inspired oxygen tension equals fraction × (barometric pressure minus water vapour pressure); falls at altitude.
- Henry: dissolved gas is proportional to partial pressure at a given temperature.
- Application: dissolved oxygen 0.003 mL/100 mL/mmHg; hyperbaric oxygen raises dissolved content; nitrogen loading in diving.
- 3.
What determines flow through a tube?
Show answer
- Laminar flow: Hagen-Poiseuille, proportional to pressure gradient and radius to the fourth power, inversely to viscosity and length.
- Turbulent flow: proportional to the square root of the pressure gradient, dependent on density.
- Reynolds number = density × velocity × diameter ÷ viscosity; above about 2000 turbulence is likely.
- Clinical: short wide cannulae for rapid infusion; small airway narrowing raises resistance steeply; heliox in large-airway obstruction.
- 4.
Explain the Bernoulli principle and the Venturi effect.
Show answer
- Conservation of energy along a streamline: pressure plus kinetic energy is constant.
- At a constriction velocity rises, so lateral pressure falls.
- The Venturi uses this low pressure to entrain a second gas in a fixed ratio.
- Uses: Venturi oxygen masks, nebulisers, suction, jet injectors.
- 5.
How does a rotameter work, and what are its sources of error?
Show answer
- Variable orifice, constant pressure difference across the bobbin.
- Low flow: narrow annulus, laminar, viscosity-dependent; high flow: orifice-like, density-dependent.
- Calibrated for one gas at a set temperature and pressure.
- Errors: static electricity, dirt, tilted tube, back pressure, a sticking bobbin, cracked tube causing leak.
- 6.
Compare variable and fixed performance oxygen devices.
Show answer
- The deciding factor is whether device flow meets the patient's peak inspiratory flow.
- Variable: nasal cannulae, simple mask, reservoir mask; fraction falls as breathing becomes faster and deeper.
- Fixed: Venturi masks (entrainment ratio fixed), ventilators and sealed circuits.
- Choice depends on the need for a precise fraction, for example in chronic hypercapnia.
- 7.
How does high-flow nasal oxygen improve gas exchange?
Show answer
- Flow high enough to meet inspiratory demand, so the oxygen fraction is predictable.
- Washout of nasopharyngeal dead space, lowering rebreathed carbon dioxide and work of breathing.
- Small flow-dependent positive airway pressure with the mouth closed, aiding recruitment.
- Heated humidified gas improves mucociliary clearance and comfort and reduces the energy cost of conditioning gas.
- 8.
Why does a nitrous oxide cylinder frost during use, and what happens to its pressure?
Show answer
- Liquid evaporates to replace vapour drawn off; latent heat of vaporisation is taken from the liquid and cylinder.
- The liquid cools, so its saturated vapour pressure falls and the gauge reading drops.
- Water vapour from the air condenses and freezes on the outside.
- Pressure recovers as the cylinder warms; only when the liquid is gone does pressure fall steadily.
- 9.
Why can an Entonox cylinder deliver a hypoxic mixture?
Show answer
- The 50:50 mixture behaves as a gas because of the Poynting effect, with a pseudocritical temperature near minus 6 °C.
- Below it, nitrous oxide liquefies and the phases separate.
- Oxygen-rich gas is delivered first, then a nitrous oxide-rich hypoxic mixture.
- Prevented by storing cylinders warm and horizontal and inverting them before use.
Practise this week
- Ventilator Waveform Reader
- First Part MCQs: measurement and equipment