Ultrasound and Doppler
The physics of ultrasound: frequency, wavelength and the speed of sound, the piezoelectric effect, acoustic impedance and reflection, attenuation, resolution and the trade-off with depth, image modes and common artefacts, the Doppler effect and its uses for measuring velocity and flow, and the safe use of ultrasound.
By the end of this page you should be able to
- Define ultrasound and relate frequency, wavelength and the speed of sound in tissue.
- Explain how a piezoelectric transducer generates and receives ultrasound and how a pulse-echo image is formed.
- Explain acoustic impedance, reflection, attenuation and the trade-off between resolution and depth.
- Recognise common artefacts and explain their causes.
- State the Doppler equation, explain the effect of the angle of insonation, and describe pulsed, continuous wave and colour Doppler and their limits.
Sound waves
Ultrasound is sound above the range of human hearing (above 20 kHz); medical imaging uses roughly 2 to 15 MHz. Sound is a longitudinal pressure wave that needs a medium. Its speed (c) depends on the medium's stiffness and density, not on frequency: about 330 m/s in air, 1540 m/s in soft tissue (the value scanners assume), about 1480 m/s in water and fat, and about 4000 m/s in bone. Speed = frequency × wavelength, so at 1540 m/s a 5 MHz wave has a wavelength of about 0.3 mm. Shorter wavelengths can resolve smaller structures.
Making and receiving the beam
Ultrasound is generated and detected by the piezoelectric effect: certain crystals or ceramics (such as lead zirconate titanate) deform when a voltage is applied, and generate a voltage when they are deformed by a returning wave. A transducer therefore acts as a loudspeaker and then a microphone. It emits a short pulse of a few cycles, then listens. A backing layer damps the crystal so pulses are short, and a matching layer and coupling gel reduce the impedance mismatch at the skin so that sound enters the body rather than reflecting from the surface.
Pulse-echo principle: the time for an echo to return gives the depth of the reflector: depth = c × time ÷ 2 (out and back). The amplitude of the echo sets brightness. Repeating this along many scan lines builds a two-dimensional image. Array transducers fire groups of elements in sequence and use timing delays to steer and focus the beam electronically. Linear arrays (high frequency, rectangular image) suit superficial structures such as vessels and nerves; curvilinear arrays (lower frequency, wide sector) suit the abdomen; phased arrays (small footprint, steered sector) suit the heart between ribs.
Reflection, attenuation and resolution
Acoustic impedance (Z) is density × speed of sound. Sound is reflected at an interface between tissues of different impedance, and the larger the mismatch, the stronger the reflection. Soft tissues differ slightly, giving the grey-scale image. Tissue and bone or gas differ hugely, so nearly all the sound is reflected and little passes beyond: hence acoustic shadowing behind bone and stones, and why air between probe and skin must be excluded with gel. Smooth surfaces larger than the wavelength act as specular reflectors and reflect best when the beam strikes them at 90°; smaller or rough structures scatter sound in all directions, giving tissue texture. Refraction occurs when sound crosses an interface obliquely between tissues with different speeds.
Attenuation is the loss of intensity with depth, from absorption (conversion to heat), reflection and scatter, measured in decibels. In soft tissue it is roughly 0.5 to 1 dB per cm per MHz, so it rises with frequency. The scanner compensates with time gain compensation, amplifying later echoes more.
This is the central trade-off: higher frequency gives better resolution but less penetration. Axial resolution (along the beam) is about half the spatial pulse length, so it improves with higher frequency and shorter pulses. Lateral resolution (across the beam) depends on beam width and is best at the focal zone. Temporal resolution depends on frame rate, which falls with greater depth, wider sectors and more focal zones.
Artefacts
- Acoustic shadowing: a dark band behind a strong reflector or attenuator (bone, calculi, gas).
- Posterior acoustic enhancement: a bright band behind a fluid-filled structure, which attenuates less than surrounding tissue while the time gain compensation still boosts it.
- Reverberation: repeated equally spaced echoes between two strong parallel reflectors (a needle, the pleura); A-lines in lung ultrasound are reverberations from the pleura, and comet tail or ring-down artefacts arise from small strong reflectors.
- Mirror image: a duplicate structure appears deep to a strongly reflective curved surface such as the diaphragm.
- Side lobe and beam width artefacts: echoes from outside the main beam are placed on the scan line, filling fluid spaces with spurious echoes.
- Anisotropy: tendons and nerves change brightness with the angle of the beam, which matters for regional anaesthesia.
- Speed artefacts: the scanner assumes 1540 m/s, so regions with different speeds misplace depth; a needle can appear bent.
Doppler
The Doppler effect is the change in frequency of a wave reflected from a moving object: higher when the reflector moves towards the probe, lower when it moves away. For red cells, the shift is given by Δf = 2 × f₀ × v × cos θ ÷ c, where f₀ is the transmitted frequency, v the velocity, θ the angle between the beam and the direction of flow, and c the speed of sound. Rearranged, v = Δf × c ÷ (2 × f₀ × cos θ). The shifts are in the audible range, which is why Doppler signals can be heard.
Angle matters: cos 0° = 1, so velocity is measured best with the beam parallel to flow; cos 90° = 0, so no shift is detected when the beam is perpendicular; and above about 60° small errors in angle produce large errors in velocity. In echocardiography the angle is assumed to be near zero, so measured velocities are always underestimates if it is not.
Continuous wave Doppler uses separate transmitting and receiving crystals; it can measure very high velocities (stenotic jets) but cannot tell at what depth along the beam they arise (range ambiguity). Pulsed wave Doppler samples a chosen depth (the sample volume) by timing, but the maximum measurable frequency shift is half the pulse repetition frequency (the Nyquist limit); above it, aliasing wraps the signal round so high velocities appear in the wrong direction. Deeper sampling lowers the pulse repetition frequency and the limit. Colour flow Doppler applies pulsed Doppler to many sample points and encodes mean velocity and direction in colour (conventionally red towards the probe and blue away), at the cost of frame rate and with the same aliasing. Tissue Doppler measures the low-velocity, high-amplitude motion of myocardium.
Velocity becomes flow when the velocity-time integral (the area under the spectral trace, or stroke distance) is multiplied by the cross-sectional area of the vessel or valve: the basis of stroke volume measurement by echocardiography and oesophageal Doppler. The simplified Bernoulli equation, pressure gradient = 4v², estimates gradients across stenoses and pulmonary artery pressure from a tricuspid regurgitant jet.
Safety
Ultrasound deposits energy as heat and causes mechanical effects (cavitation and radiation force). Scanners display a thermal index (the ratio of power used to the power needed to raise tissue temperature by 1 °C) and a mechanical index (related to peak negative pressure and inversely to the square root of frequency). Doppler, especially pulsed Doppler held in one place, deposits more energy than imaging. The principle is to use the lowest output and shortest time needed (ALARA), with particular care for the fetus, the eye and in the presence of gas contrast agents. Probe hygiene and sterile covers for procedures prevent infection.
What examiners reward
- Start with the wave equation and the speed in tissue, then the piezoelectric effect, then pulse-echo.
- Explain resolution and penetration as one trade-off controlled by frequency.
- Write the Doppler equation and explain the cos θ term with a sketch.
- Contrast continuous and pulsed Doppler by range ambiguity and aliasing.
Common traps
- Saying higher frequency penetrates further; it attenuates more.
- Saying reflection depends on density alone; it depends on the difference in acoustic impedance.
- Measuring velocity with the beam perpendicular to flow.
- Confusing aliasing (pulsed Doppler, Nyquist limit) with range ambiguity (continuous wave Doppler).
- Forgetting the factor of two in both the depth and Doppler equations (the round trip).
Where this meets anaesthesia
- Use a high-frequency linear probe for superficial vessels and nerves; it resolves more detail but penetrates less.
- Keep the needle in plane and look for reverberation artefact to confirm the tip.
Viva questions
Answer each one aloud, as you would to an examiner, before you open the model points.
- 1.
What is ultrasound and how is it produced?
Show answer
- Longitudinal pressure waves above 20 kHz; medical imaging roughly 2 to 15 MHz.
- Piezoelectric crystals deform when a voltage is applied and generate a voltage when deformed.
- Short pulses emitted, echoes received by the same crystal.
- Backing layer shortens pulses; matching layer and gel reduce reflection at the skin.
- 2.
How is an ultrasound image formed?
Show answer
- Pulse-echo: depth = speed × time ÷ 2, assuming 1540 m/s.
- Echo amplitude sets brightness; many scan lines build the image.
- Reflection at interfaces of differing acoustic impedance; scatter gives texture.
- Time gain compensation corrects for attenuation with depth.
- 3.
Why can you not image through bone or bowel gas?
Show answer
- Acoustic impedance of bone and gas differs greatly from soft tissue.
- Almost all the sound is reflected at the interface, leaving little to travel further.
- Produces acoustic shadowing beyond.
- Same reason gel is needed between probe and skin.
- 4.
Explain the relationship between frequency, resolution and penetration.
Show answer
- Higher frequency, shorter wavelength and pulse: better axial resolution.
- Attenuation rises with frequency, so penetration falls.
- High frequency linear probes for superficial nerves and vessels; low frequency for deep structures.
- Lateral resolution depends on beam width and focus; temporal resolution on frame rate.
- 5.
Describe some ultrasound artefacts and their causes.
Show answer
- Shadowing behind strong reflectors; enhancement behind fluid.
- Reverberation between parallel strong reflectors, including A-lines and needle reverberation.
- Mirror image deep to the diaphragm.
- Side lobe and beam width artefacts, anisotropy, speed displacement artefacts.
- 6.
State the Doppler equation and explain the importance of the angle.
Show answer
- Δf = 2 f₀ v cos θ ÷ c.
- Beam parallel to flow gives the full shift; perpendicular gives none.
- Above about 60° small angle errors give large velocity errors.
- Assuming zero angle when it is not underestimates velocity.
- 7.
Compare continuous wave and pulsed wave Doppler.
Show answer
- Continuous: separate send and receive crystals; measures high velocities; cannot localise depth.
- Pulsed: samples a chosen depth by timing; limited by the Nyquist limit of half the pulse repetition frequency.
- Above the limit, aliasing; deeper sampling lowers the limit.
- Colour Doppler is pulsed Doppler across many points, showing mean velocity and direction.
- 8.
How can Doppler be used to measure cardiac output?
Show answer
- Velocity-time integral of flow in the outflow tract or aorta gives stroke distance.
- Multiplied by cross-sectional area gives stroke volume; times heart rate gives output.
- Errors from the angle of insonation and from estimating the area, which is squared from a diameter.
- Used by transthoracic and transoesophageal echocardiography and oesophageal Doppler probes.
- 9.
Is ultrasound safe?
Show answer
- Thermal effects from absorption; mechanical effects including cavitation.
- Thermal and mechanical indices displayed to guide output.
- Doppler, especially pulsed Doppler at one site, deposits more energy.
- Use the lowest output for the shortest time; take care with the fetus, the eye and contrast agents.
Practise this week
- Primary MCQs: measurement and equipment
- Primary MCQs: quality and safety
- Primary SAQs, marked part by part
- Primary vivas