Neuromuscular and depth of anaesthesia monitoring
How a peripheral nerve stimulator works, the patterns of stimulation (single twitch, train-of-four, tetanus, post-tetanic count and double burst) and what they show for depolarising and non-depolarising block, quantitative monitoring and residual block, the physiological basis of the EEG, how anaesthetics change it, and the principles and limits of processed EEG and evoked potential monitoring.
By the end of this page you should be able to
- Describe the principles of peripheral nerve stimulation, including supramaximal stimulation and the choice of nerve and muscle.
- Describe the patterns of stimulation and the responses seen in depolarising, non-depolarising and phase II block.
- Explain why qualitative assessment misses residual block and how quantitative monitors measure it.
- Describe the physiological basis of the EEG and how it changes with anaesthesia.
- Explain the principles and limitations of processed EEG monitors and outline evoked potentials.
Stimulating the nerve
A peripheral nerve stimulator delivers a brief square-wave current (pulse width about 0.1 to 0.3 ms) through surface electrodes over a nerve, and the response of the muscle it supplies is observed or measured. The stimulus must be supramaximal, about 10 to 20% above the current that recruits every fibre (typically 50 to 60 mA in an adult), so that any fall in response reflects neuromuscular block rather than incomplete stimulation. Because skin impedance varies, the device is a constant current stimulator. The negative (black) electrode is placed distally over the nerve.
The usual site is the ulnar nerve at the wrist, with the adductor pollicis as the responding muscle; it reflects recovery of upper airway and peripheral muscles. Other sites are the facial nerve (corrugator supercilii, which behaves more like the laryngeal and diaphragmatic muscles and recovers earlier) and the posterior tibial nerve. Different muscles have different sensitivity: the diaphragm and laryngeal muscles are relatively resistant (block sets in and wears off faster), while adductor pollicis and upper airway dilator muscles are more sensitive. Monitoring a resistant muscle can overestimate recovery of the airway.
Patterns of stimulation
- Single twitch (often at 0.1 Hz or 1 Hz): needs a control height before the drug; useful for onset only.
- Train-of-four (TOF): four stimuli at 2 Hz, repeated no more often than every 10 to 15 seconds. The TOF count (number of responses) and the TOF ratio (fourth twitch ÷ first twitch) need no control. With non-depolarising block, the twitches fade because the drug also blocks presynaptic nicotinic receptors that normally mobilise acetylcholine during repeated stimulation; the fourth twitch disappears at roughly 75% receptor occupancy and the first at about 90 to 95%. With depolarising block (suxamethonium, phase I), all four responses are reduced equally without fade.
- Tetanic stimulation (usually 50 Hz for 5 seconds): sustained contraction in normal function; fade with non-depolarising block; no fade in phase I depolarising block. It is painful and can alter subsequent responses at that site for several minutes.
- Post-tetanic count (PTC): after a tetanus, single twitches at 1 Hz are counted. The tetanus mobilises acetylcholine, so for a short time post-tetanic potentiation produces responses even during deep block when the TOF count is zero. The fewer the post-tetanic twitches, the deeper the block and the longer until the first TOF response returns. Useful to quantify deep block and to judge reversal of rocuronium with sugammadex.
- Double burst stimulation (DBS): two short bursts of 50 Hz stimuli 750 ms apart; fade between the two bursts is easier to feel than TOF fade, but still misses residual block at higher ratios.
- Phase II block: after prolonged or repeated suxamethonium, or with reduced plasma cholinesterase, the depolarising block takes on non-depolarising characteristics: fade on TOF and tetanus, and post-tetanic potentiation.
Residual block and quantitative monitoring
Clinical tests (head lift, grip, tidal volume) and tactile or visual assessment of fade are unreliable: observers cannot reliably detect fade once the TOF ratio exceeds about 0.4, yet a ratio below 0.9 is associated with impaired pharyngeal function, aspiration, airway obstruction, reduced hypoxic ventilatory response and postoperative pulmonary complications. Adequate recovery is a measured TOF ratio of at least 0.9 at the adductor pollicis. Quantitative monitors measure the response objectively:
Acceleromyography measures acceleration of the thumb with a piezoelectric sensor (force = mass × acceleration, so acceleration is proportional to force); it is the most common, but needs a free-moving thumb and a baseline, and it tends to over-read the TOF ratio (normalising to a baseline ratio helps). Electromyography records the compound muscle action potential, needs no free movement, works with the arm tucked, and is less affected by position. Kinemyography measures bending of a piezoelectric strip between thumb and finger. Mechanomyography measures isometric force with a transducer and is the research reference standard.
The link with pharmacology: reversal with an anticholinesterase works only once there is some spontaneous recovery (a TOF count of at least two to four, depending on the drug and the goal), because it raises acetylcholine to compete and has a ceiling effect; sugammadex encapsulates rocuronium and vecuronium and can reverse deep block, with the amount guided by the depth measured on TOF or PTC (details belong to local guidelines and the drug information).
The EEG and how anaesthesia changes it
The EEG records summed postsynaptic potentials from large populations of cortical pyramidal neurones, aligned perpendicular to the surface, through the skull and scalp, as microvolt signals. Rhythms are generated by interaction between the thalamus and cortex. Frequency bands: delta (below 4 Hz, deep sleep and deep anaesthesia), theta (4 to 8 Hz), alpha (8 to 13 Hz, relaxed wakefulness with eyes closed), beta (13 to 30 Hz, alert) and gamma (above 30 Hz).
Most hypnotics acting at GABA-A receptors (propofol, volatile agents) produce a predictable sequence: an initial paradoxical excitation with increased beta, then anteriorisation of alpha with strong frontal alpha and slow delta oscillations during general anaesthesia, then burst suppression (bursts separated by flat periods) at deeper levels, and finally an isoelectric trace. Ketamine, nitrous oxide and xenon (NMDA antagonists) behave differently, increasing high-frequency activity. Hypothermia, hypoglycaemia, hypoxia and cerebral hypoperfusion also slow or suppress the EEG.
Processed EEG and evoked potentials
Processed EEG monitors record from frontal electrodes, digitise the signal, remove artefact, and analyse it in the time domain (amplitude, burst suppression ratio) and frequency domain (by Fourier transformation, giving a power spectrum; measures such as the spectral edge frequency, the frequency below which most of the power lies). Proprietary algorithms combine these into a dimensionless index, usually from about 100 awake towards 0 isoelectric, with a target range for general anaesthesia. Other approaches use entropy (the irregularity of the signal, with a separate fast component reflecting frontal muscle activity) or display the density spectral array so that the pattern itself can be read.
Uses: titrating hypnotic in total intravenous anaesthesia with neuromuscular block, where movement cannot warn of light anaesthesia, and avoiding excessive depth, especially in the elderly. Limitations: the index lags by seconds to a minute; it is distorted by EMG from frontal muscles, diathermy, pacemakers and movement; it is not validated for ketamine, nitrous oxide or xenon and is affected by hypothermia, hypoglycaemia, brain injury, dementia and age (older patients have lower amplitude EEG); a low index does not guarantee absence of awareness, and evidence that monitoring reduces awareness compared with end-tidal agent monitoring is mixed.
Evoked potentials average the EEG response to repeated stimuli to pull a small signal out of background noise: somatosensory (stimulating a peripheral nerve and recording over the cortex, for spinal and cerebrovascular surgery), motor (transcranial stimulation, recording from muscles; incompatible with deep neuromuscular block and sensitive to volatile agents), auditory (brainstem and middle latency responses, the latter changing with anaesthetic depth) and visual. Changes in latency and amplitude warn of neural compromise; volatile agents and nitrous oxide depress cortical responses more than intravenous techniques do.
What examiners reward
- Define supramaximal stimulation and explain why constant current is used before describing patterns.
- For each pattern, give the stimulus parameters, then the response in non-depolarising, depolarising phase I and phase II block, and explain fade by presynaptic receptors.
- State the TOF ratio threshold for adequate recovery and why visual assessment cannot confirm it.
- For processed EEG, describe the signal, the processing steps, and then the limitations.
Common traps
- Saying fade is seen in phase I suxamethonium block.
- Saying four visible twitches without obvious fade means full recovery.
- Saying the diaphragm is more sensitive than adductor pollicis; it is more resistant.
- Treating a processed EEG index as a direct measure of awareness.
- Applying tetanus repeatedly at the same site and then trusting the next TOF.
Where this meets anaesthesia
- Visual or tactile assessment cannot detect residual block once the TOF ratio exceeds about 0.4; use quantitative monitoring.
- Processed EEG monitors are not reliable with ketamine or nitrous oxide.
Viva questions
Answer each one aloud, as you would to an examiner, before you open the model points.
- 1.
What is supramaximal stimulation, and why is it needed?
Show answer
- A current about 10 to 20% above that which recruits all motor fibres in the nerve.
- Ensures any reduction in response reflects block, not incomplete stimulation.
- Constant current stimulators compensate for changing skin impedance.
- Short square-wave pulses avoid repetitive firing.
- 2.
Describe the train-of-four and the responses in different types of block.
Show answer
- Four stimuli at 2 Hz; count and ratio of fourth to first; no control needed.
- Non-depolarising: progressive fade from presynaptic receptor block; count falls as block deepens.
- Depolarising phase I: equal reduction, no fade.
- Phase II: fade like non-depolarising block.
- 3.
What is the post-tetanic count and when would you use it?
Show answer
- Tetanus at 50 Hz for 5 seconds, then single twitches at 1 Hz counted.
- Post-tetanic potentiation from mobilised acetylcholine gives responses during deep block.
- Fewer responses mean deeper block and longer to first TOF response.
- Used when the TOF count is zero, for deep block and to guide reversal.
- 4.
Why does fade occur with non-depolarising block?
Show answer
- Presynaptic nicotinic receptors normally provide positive feedback to mobilise acetylcholine during repetitive stimulation.
- Non-depolarising drugs block these as well as the postjunctional receptors.
- Release falls with each successive stimulus while postjunctional margin is reduced.
- Suxamethonium in phase I does not produce this.
- 5.
How would you know that neuromuscular block has fully recovered?
Show answer
- Quantitative TOF ratio of at least 0.9 at the adductor pollicis.
- Tactile or visual assessment cannot detect fade above about 0.4.
- Clinical tests such as head lift are insensitive.
- Residual block impairs pharyngeal function and hypoxic ventilatory response.
- 6.
Compare acceleromyography and electromyography.
Show answer
- Acceleromyography: piezoelectric sensor measures thumb acceleration, proportional to force; needs free movement, baseline normalisation, may over-read.
- Electromyography: compound muscle action potential; no movement needed, works with arms tucked.
- Mechanomyography is the reference but impractical clinically.
- All need supramaximal, consistent stimulation and temperature attention.
- 7.
What does the EEG record and how does propofol change it?
Show answer
- Summed postsynaptic potentials of cortical pyramidal neurones; thalamocortical rhythms.
- Bands: delta, theta, alpha, beta, gamma.
- Propofol: initial beta excitation, then frontal alpha and slow delta oscillations, then burst suppression and isoelectricity.
- NMDA antagonists produce different patterns.
- 8.
How does a processed EEG monitor work and what are its limitations?
Show answer
- Frontal electrodes; digitisation, artefact rejection; time and frequency domain analysis including Fourier power spectrum and burst suppression ratio.
- Algorithm combines measures into an index from awake towards isoelectric.
- Limitations: lag, EMG and electrical interference, not valid for ketamine or nitrous oxide, affected by age, temperature, glucose and brain pathology.
- Does not directly measure awareness.
- 9.
What are evoked potentials and how do anaesthetics affect them?
Show answer
- Averaged EEG or muscle responses to repeated stimuli, extracting small signals from noise.
- Somatosensory, motor, auditory and visual; latency and amplitude monitored.
- Volatile agents and nitrous oxide depress cortical responses more than intravenous agents.
- Motor evoked potentials need little or no neuromuscular block.
Practise this week
- Primary MCQs: measurement and equipment
- Primary MCQs: quality and safety
- Primary SAQs, marked part by part
- Primary vivas