RANZCR Clinical Radiology Phase 1 Written (Anatomy and Applied Imaging Technology) 2024 Sitting 2: what the examiners said
Report released January 2025 · Summary checked
The headline
Outdated beliefs about CT pitch and dose, and weak reading of unfamiliar views such as oblique spine and pelvic ultrasound, drove the low anatomy pass rate.
An independent plain-English summary. Fellowship Ready is not affiliated with or endorsed by the RANZCR. Read the original report on the RANZCR website. Read the full report.
Format: AIT: 60 MCQs and 9 constructed-response questions. Anatomy: 120 labels, 60 MCQs, 30 very short answers and 20 short answers across six regions. Both 3 hours with online proctoring.
What candidates did well
Explaining radiation risk to a worried patient and the use of DRLs were well done.
Fluoroscopy protection practice, X-ray production basics and window width and level were sound.
Labelling was the best anatomy format, with strong answers on cavernous sinus connections and the axillary artery.
Where marks were lost, and how to fix it
CT pitch and dose with AEC
What went wrong: Many still believed increasing pitch reduces dose, which is not true on modern scanners that hold effective mAs constant.
How to fix it: Learn how effective mAs and tube current modulation interact with pitch on current scanners.
Fluoroscopy magnification and binning
What went wrong: The dose and resolution effects of magnification modes and detector binning were poorly understood.
How to fix it: Work through what happens to field size, dose rate, resolution and noise when the field of view is reduced or pixels are binned.
Bit depth and display
What went wrong: Some thought window level changes spatial resolution or that CT data are 8-bit.
How to fix it: Relate stored bit depth to the HU range and explain windowing as mapping a chosen range onto display grey levels.
MR implant labelling
What went wrong: Implant categories were often wrong, with the outdated term 'MR compatible' used.
How to fix it: Use only MR Safe, MR Conditional and MR Unsafe, and know what each means for scanning.
Ultrasound frequency terms
What went wrong: Transmit frequency was linked to the wrong property, and PRF was not separated from transmit frequency.
How to fix it: Tie transmit frequency to attenuation and axial resolution, and PRF to depth and aliasing, as separate ideas.
Unfamiliar imaging views
What went wrong: Oblique lumbar spine, longitudinal pelvic ultrasound and axial ankle MRI were the weakest images.
How to fix it: Deliberately practise labelling less common projections and planes, not only standard axial CT.
Embryology-based variants and branches
What went wrong: Pancreatic duct variants, renal segmental arteries, first arch muscles and internal pudendal branches were poorly answered.
How to fix it: Learn variants from their embryological origin, and learn vessels down to named branch level.
Advice for the next sitting
Read the whole question and identify exactly what is being asked before writing.
Check that your physics reflects current scanner behaviour, not older teaching.
Always include units and check them.
In two-part comparisons, answer both halves.
Practise unfamiliar views and planes for every region.
Original questions written to train the themes above (not questions from the exam). Answer each to see the explanation and why the other options are wrong.
Question 1 of 5CT pitch and dose with AEC
On a modern CT scanner whose dose is set by effective mAs (mAs divided by pitch) with automatic exposure control, the pitch for a chest protocol is increased from 1.0 to 1.5. What happens to patient dose?
Pick an answer to see the explanation.
Fix what the examiners said
The same habits cost marks sitting after sitting. Each one has a short drill of questions from the Radiology bank that test exactly that habit.