Medical gases, the anaesthetic machine and vaporisers
How medical gases are produced, stored and piped, the safety features of cylinders and pipelines, the path of gas through the anaesthetic machine from supply to common gas outlet, the machine's mandatory safety features, and the physics and design of vaporisers, including the factors that change their output and the special case of desflurane.
By the end of this page you should be able to
- Describe how oxygen, nitrous oxide and medical air are produced, stored and distributed, and the safety features of cylinders and pipelines.
- Trace the path of gas through an anaesthetic machine and explain the purpose of each component.
- List the safety features designed to prevent delivery of a hypoxic mixture and to protect against supply failure.
- Explain the principles of the variable bypass plenum vaporiser and the factors that affect its output.
- Explain why desflurane needs a different vaporiser and outline how it works.
Medical gas supply
Oxygen is produced by fractional distillation of liquefied air, stored in hospitals as a liquid in a vacuum-insulated evaporator (a double-walled flask kept below oxygen's critical temperature, with liquid evaporating to supply the pipeline; heat taken from the surroundings boils it, and a superheater warms it before use). Smaller sites and reserves use banks of cylinders, and some sites use oxygen concentrators, which pass air through zeolite that adsorbs nitrogen (pressure swing adsorption), producing about 90 to 96% oxygen with argon as the main remainder.
Nitrous oxide is made by heating ammonium nitrate, purified to remove higher oxides of nitrogen, and stored as a liquid in cylinders. Medical air is compressed, filtered and dried from atmospheric air.
Piped gases reach the wall at about 400 kPa (4 bar) and are delivered through non-interchangeable terminal outlets that accept only the matching probe, with colour coding and labelling. Cylinders connect to machines through a pin index system (pins on the yoke matching holes on the valve block, so a cylinder of one gas cannot be fitted to another gas's yoke) and have a Bodok seal. Cylinders are colour coded and labelled by gas under the relevant Australian and New Zealand standards. Gas from cylinders passes through a pressure regulator that reduces it to a working pressure similar to the pipeline.
Through the machine
- High pressure system: cylinder supply (with yoke, gauge and regulator) and pipeline inlets with non-return valves and gauges. The machine preferentially draws from the pipeline because the cylinder regulator is set slightly lower.
- Intermediate pressure system: pipeline pressure gas feeds the oxygen failure warning, the oxygen flush (delivering a high flow of pure oxygen directly to the common gas outlet, bypassing the flowmeters and vaporiser), the ventilator driving gas, and the flow control valves. Secondary regulators may smooth pipeline fluctuations.
- Low pressure system: downstream of the flow control valves: flowmeters (rotameters or electronic sensors), the back bar with the vaporisers, a non-return valve and pressure relief valve, and the common gas outlet to the breathing system.
- Flowmeters: the oxygen flowmeter is placed last (downstream) in the bank, so a crack in another flowmeter tube leaks other gases rather than oxygen. Rotameters are gas-specific and calibrated for their own gas (see the gas laws page for how they work).
- Electronic machines replace rotameters with flow sensors and electronic mixers, but keep pneumatic backups and a means of delivering oxygen if power fails.
Safety features
- Hypoxic guard (proportioning system): mechanical or electronic linkage between the oxygen and nitrous oxide controls that prevents a nitrous oxide to oxygen mixture with less than about 25% oxygen. It does not protect against a hypoxic mixture with a third gas or a leak downstream, so inspired oxygen monitoring is still essential.
- Oxygen failure warning: an alarm, powered by the oxygen supply pressure itself, sounds when supply pressure falls; nitrous oxide is cut off automatically (and air may be allowed), so the machine fails to air or to nothing rather than to hypoxic nitrous oxide.
- Minimum oxygen flow: many machines deliver a basal oxygen flow whenever they are on.
- Pin index, non-interchangeable connectors, colour coding and gas-specific flowmeter knobs (the oxygen knob is larger and fluted).
- Vaporiser interlock: only one vaporiser can be turned on at a time, preventing mixing of agents.
- Pressure relief on the back bar protects the machine; the adjustable pressure limiting valve and ventilator alarms protect the patient.
- Monitoring and checks: inspired oxygen, airway pressure and capnography, alarms, and a documented machine check before use; ANZCA professional documents set the mandatory standards for anaesthesia delivery systems.
Vaporisers: the physics
A volatile agent in a closed container evaporates until the vapour above the liquid reaches its saturated vapour pressure (SVP), which depends only on the agent and its temperature. Any gas that passes over the liquid long enough leaves saturated: its concentration is SVP ÷ ambient pressure. For sevoflurane at 20 °C the SVP is about 160 mmHg, so saturated gas is about 21%; for isoflurane about 240 mmHg (about 32%). These are far above clinical concentrations, so most of the fresh gas must bypass the agent.
A variable bypass plenum vaporiser splits the fresh gas into a bypass stream and a smaller stream through the vaporising chamber, where wicks greatly increase the surface area so the gas leaves fully saturated. The two streams rejoin, and the splitting ratio, set by the control dial, determines the output. Because the chamber gas is saturated, output depends on SVP and the splitting ratio.
The design keeps those two constant as conditions change:
Temperature: evaporation takes latent heat from the liquid, so it cools during use and its SVP falls. This is opposed by building the vaporiser from metal with a high specific heat capacity (a heat reservoir) and high thermal conductivity (drawing heat from the room), and by temperature compensation: a bimetallic strip or an expanding capsule that lets more gas into the chamber as the temperature falls.
Fresh gas flow: output is fairly constant over the usual range; at very low flows there is too little turbulence to carry vapour out, and at very high flows the gas may not be fully saturated, so output falls slightly at both extremes.
Back pressure (the pumping effect): intermittent pressure from a ventilator or the oxygen flush can push gas back into the chamber and then out through the bypass, raising output. Long inlet tubes, baffles and non-return valves prevent it.
Barometric pressure: a plenum vaporiser delivers a roughly constant partial pressure of agent, which is what matters clinically, so at altitude the dial percentage under-reads the concentration but the effect is similar. Carrier gas composition changes output slightly, transiently with nitrous oxide as it dissolves in the agent.
Agent-specific filling systems (keyed fillers) prevent filling with the wrong agent; a wrong agent would deliver a concentration set by its own SVP.
Desflurane and other designs
Desflurane has an SVP of about 670 mmHg at 20 °C and boils at about 23 °C, so a small change in room temperature would cause a large change in output, and in a plenum vaporiser it could boil and deliver uncontrollably high concentrations. Its vaporiser heats the agent to about 39 °C in a sump, giving a stable SVP of about 2 atmospheres. The pressurised vapour is then injected into the fresh gas in proportion to fresh gas flow by an electronically controlled pressure-regulating valve, so it is a dual-gas blender rather than a bypass device. It needs electrical power, and because it delivers a set concentration (not partial pressure), its effect falls at altitude and the setting must be increased.
Other designs include injector vaporisers (electronic machines that inject measured liquid agent into the gas), the draw-over vaporiser (low resistance, used with a self-inflating bag where compressed gas is unavailable), and the old copper kettle (measured flow vaporiser).
What examiners reward
- Trace gas from supply to patient, naming high, intermediate and low pressure parts, and give each safety feature its place.
- For vaporisers, start from SVP and full saturation, then the splitting ratio, then each factor that would change output and the design feature that counters it.
- Explain desflurane from its boiling point and SVP, then the heated, pressurised blender.
- Know which safety features protect against which failure, and which do not.
Common traps
- Saying the hypoxic guard makes a hypoxic mixture impossible; it only links oxygen and nitrous oxide.
- Putting the oxygen flowmeter first in the bank.
- Saying a plenum vaporiser delivers a constant percentage at altitude; it delivers a nearly constant partial pressure.
- Forgetting that the oxygen flush bypasses the vaporiser and can cause barotrauma or awareness.
- Saying desflurane is vaporised by drawing gas over wicks.
Where this meets anaesthesia
- Inspired oxygen monitoring is the final safeguard against a hypoxic mixture, because the hypoxic guard only links oxygen to nitrous oxide.
- Filling a vaporiser with the wrong agent changes the output according to that agent's saturated vapour pressure.
Viva questions
Answer each one aloud, as you would to an examiner, before you open the model points.
- 1.
How is oxygen supplied to a hospital?
Show answer
- Fractional distillation of liquefied air; stored as liquid in a vacuum-insulated evaporator.
- Evaporation supplies the pipeline at about 400 kPa through a superheater and regulators.
- Reserve cylinder banks; oxygen concentrators using zeolite pressure swing adsorption in some settings.
- Non-interchangeable outlets, colour coding and alarms protect the system.
- 2.
Trace the path of oxygen through an anaesthetic machine.
Show answer
- Pipeline or cylinder (via regulator) into the high pressure system with gauges and non-return valves.
- Intermediate pressure: oxygen failure alarm, flush, ventilator driving gas, flow control valve.
- Low pressure: flowmeter (last in the bank), back bar, vaporiser, non-return and relief valves.
- Common gas outlet to the breathing system.
- 3.
What features of the machine prevent a hypoxic mixture?
Show answer
- Hypoxic guard linking nitrous oxide to oxygen flow.
- Oxygen failure warning and nitrous oxide cut-off.
- Oxygen flowmeter downstream; minimum oxygen flow; pin index and non-interchangeable connectors.
- Inspired oxygen monitoring, because none of the others covers every failure.
- 4.
Explain how a plenum vaporiser works.
Show answer
- Fresh gas split into bypass and chamber streams by the dial.
- Wicks ensure gas leaving the chamber is saturated at the agent's SVP.
- Output depends on SVP and the splitting ratio.
- Temperature compensation, heat sink, and baffles keep these constant.
- 5.
What factors affect the output of a vaporiser?
Show answer
- Temperature: latent heat cools the liquid, lowering SVP; countered by metal mass, conductivity and compensation.
- Fresh gas flow: output falls a little at very low and very high flows.
- Back pressure: pumping effect raises output; prevented by baffles and valves.
- Barometric pressure, carrier gas composition, tilting and wrong agent.
- 6.
Why does desflurane need a special vaporiser?
Show answer
- SVP near atmospheric and boiling point about 23 °C, so output in a plenum would vary hugely and it could boil.
- Heated to about 39 °C to give a stable pressurised vapour.
- Vapour injected in proportion to fresh gas flow: a dual-gas blender.
- Needs power; delivers a set concentration, so the setting must rise at altitude.
- 7.
What happens at altitude to a plenum vaporiser's output?
Show answer
- SVP depends on temperature, not ambient pressure.
- Concentration rises as ambient pressure falls, but the partial pressure stays about the same.
- Clinical effect depends on partial pressure, so the effect is similar to sea level.
- Contrast with the desflurane vaporiser, which delivers a set concentration and so a lower partial pressure.
- 8.
What would happen if a sevoflurane vaporiser were filled with isoflurane?
Show answer
- Output depends on the SVP of the agent in the chamber.
- Isoflurane has a higher SVP, so more vapour is delivered than the dial suggests.
- Isoflurane is also more potent, compounding the overdose.
- Keyed filling systems prevent this; agent monitoring detects it.
- 9.
How is a cylinder made safe to connect to a machine?
Show answer
- Pin index system on the yoke matches only that gas's valve block.
- Bodok seal for a gas-tight connection.
- Colour coding and labelling; contents gauge.
- Regulator reduces pressure to working pressure; open the valve slowly, away from the face; no oil or grease.
Practise this week
- Primary MCQs: measurement and equipment
- Primary MCQs: quality and safety
- Primary SAQs, marked part by part
- Primary vivas